The answers to Part 2 of our QC basic interpretation series. Now we see what happens when we have 2 controls, and need to interpret across materials, across levels, and across runs.
We're continuing in our effort to break all the "Westgard Rules." As we'll soon see, Part One was easy - just one control to worry about. Now that we've got two controls, we're going to have to start looking in many directions to detect the errors.
Recall that low control has a mean of 47 and an SD of 3 and a high control has a mean of 71 and an SD of 8. Rather than deal in the "raw" numbers, we're going to convert those values into z-scores, which tell you how many standard deviations each point is from the mean. It's easier to pick up the rule violations this way.
Control Measurement |
Value 1
|
Z-value 1
|
Value 2 |
Z-value 2 |
Rule Broken |
| 1 | 51.5 | 1.5 | 82.2 | 1.4 | |
| 2 | 38.9 | -2.7 | 51 | -2.5 | 2:2s on both |
| 3 | 47.6 | 0.2 | 74.2 | 0.4 | |
| 4 | 46.0 | -0.33 | 91.8 | 2.6 | |
| 5 | 50.9 | 1.3 | 74.2 | 0.4 | |
| 6 | 53.9 | 2.3 | 69.0 | -0.25 | |
| 7 | 52.7 | 1.9 | 58.2 | -1.6 | |
| 8 | 49.1 | 0.7 | 95.5 | 3.06 | 1:3s on 2 |
| 9 | 48.5 | 0.5 | 65.4 | -0.7 | |
| 10 | 37.0 | -3.33 | 90.0 | 2.375 | 1:3s on 1 and R:4s on both |
| 11 | 47.0 | 0 | 63.0 | -1 | |
| 12 | 54.0 | 2.33 | 82.2 | 1.4 | |
| 13 | 53.3 | 2.1 | 65.4 | -0.7 | 2:2s on 1 |
| 14 | 40.0 | -2.33 | 86.2 | 1.9 | R:4s by calc |
| 15 | 55.4 | 2.8 | 70.2 | -0.1 | |
| 16 | 49.4 | 0.8 | 88.6 | 2.2 | |
| 17 | 46.7 | -0.1 | 87.8 | 2.1 | 2:2s on 2 |
| 18 | 45.8 | -0.4 | 85.0 | 1.75 | |
| 19 | 52.7 | 1.9 | 76.6 | 0.7 | |
| 20 | 51.1 | 1.4 | 65.4 | -0.7 | |
| 21 | 54.0 | 2.33 | 86.2 | 1.9 | |
| 22 | 53.0 | 2 | 71.0 | 0 | 4:1s on 1 and 2:2s on 1 |
| 23 | 49.7 | 0.9 | 81.4 | 1.3 | |
| 24 | 48.2 | 0.4 | 90.2 | 2.4 | |
| 25 | 46.4 | -0.2 | 80.6 | 1.2 | |
| 26 | 54.2 | 2.4 | 54.0 | -2.12 | R:4s |
| 27 | 51.2 | 1.4 | 53 | -2.25 | |
| 28 | 53.9 | 2.3 | 76.6 | 0.7 | |
| 29 | 44.3 | -0.9 | 72.6 | 0.2 | |
| 30 | 48.5 | 0.5 | 84.6 | 1.7 |
To make it a little easier to see the errors, we've marked them on a Levey-Jennings chart with z-scores for these first 30 values:
Control Measurement |
Value 1
|
Z-value 1
|
Value 2 |
Z-value 2 |
Rule Broken |
| 31 | 48.5 | 0.5 | 85.4 | 1.8 | |
| 32 | 48.8 | 0.6 | 86.2 | 1.9 | |
| 33 | 46.7 | -0.1 | 80.0 | 1.12 | |
| 34 | 46.1 | -0.3 | 79.9 | 1.11 | 4:1s on 2 |
| 35 | 46.0 | -0.33 | 83.8 | 1.6 | |
| 36 | 53.9 | 2.3 | 68.0 | -0.37 | |
| 37 | 43.0 | -1.33 | 64.6 | -0.8 | |
| 38 | 42.0 | -1.67 | 71.0 | 0 | |
| 39 | 40.0 | -2.33 | 83.0 | 1.5 | |
| 40 | 42.5 | -1.5 | 67.0 | -0.5 | 4:1s on 1 |
| 41 | 50.3 | 1.1 | 74.2 | 0.4 | |
| 42 | 44.3 | -0.9 | 46.0 | -3.12 | 1:3s on 2 |
| 43 | 51.8 | 1.6 | 71.8 | 0.1 | |
| 44 | 41.9 | -1.7 | 59.8 | -1.4 | |
| 45 | 40.0 | -2.33 | 57.0 | -1.75 | 4:1s on both |
| 46 | 48.8 | 0.6 | 86.2 | -2 | |
| 47 | 53.0 | 2 | 55.0 | -0.12 | |
| 48 | 52.4 | 1.8 | 70.0 | 1 | |
| 49 | 38.0 | -3 | 79.0 | -0.12 | 1:3s on 1 |
| 50 | 49.1 | 0.7 | 70.0 | -0.25 | |
| 51 | 44.0 | -1 | 69.0 | -0.25 | |
| 52 | 48.8 | 0.6 | 75.0 | 0.5 | |
| 53 | 50.6 | 1.2 | 76.6 | 0.7 | |
| 54 | 43.7 | -1.1 | 72.6 | 0.2 | |
| 55 | 47.6 | 0.2 | 80.0 | 1.12 | |
| 56 | 45.5 | -0.5 | 79.0 | 1 | |
| 57 | 44.3 | -0.9 | 74.0 | 0.37 | |
| 58 | 52.4 | 1.8 | 82.0 | 1.37 | |
| 59 | 53.9 | 2.3 | 72.0 | 0.12 | 8:x on 2 |
Here is the second set of Levey-Jennings charts for values 31 through 59:
Control Measurement |
Value 1
|
Z-value 1
|
Value 2 |
Z-value 2 |
Rule Broken |
| 60 | 45.8 | -0.4 | 74.2 | 0.4 | |
| 61 | 50.6 | 1.2 | 64.6 | -0.8 | |
| 62 | 48.0 | 0.33 | 71.0 | 0 | |
| 63 | 49.0 | 0.67 | 83.0 | 1.5 | |
| 64 | 51.0 | 1.33 | 67.0 | -0.5 | |
| 65 | 53.0 | 2 | 74.2 | 0.4 | |
| 66 | 48.0 | 0.33 | 80.6 | 1.2 | |
| 67 | 49.0 | 0.67 | 71.8 | 0.1 | |
| 68 | 50.0 | 1 | 69.4 | -0.2 | 8:x on 1 |
| 69 | 47.8 | 0.26 | 81.4 | 1.3 | |
| 70 | 45.5 | -0.5 | 65.4 | -0.7 | |
| 71 | 46.0 | -0.33 | 70.0 | -0.12 | |
| 72 | 45.0 | -0.67 | 69.0 | -0.25 | |
| 73 | 46.0 | -0.33 | 66.0 | -0.62 | 8:x on both |
| 74 | 47.9 | 0.3 | 68.6 | -0.3 | |
| 75 | 48.5 | 0.5 | 86.2 | 1.9 | |
| 76 | 50.9 | 1.3 | 59.8 | -1.4 | |
| 77 | 46.7 | -0.1 | 67.8 | -0.4 | |
| 78 | 47.8 | 0.26 | 86.2 | 1.9 | |
| 79 | 50.3 | 1.1 | 72.6 | 0.2 | |
| 80 | 52.0 | 1.67 | 81.4 | 1.3 | |
| 81 | 52.0 | 1.67 | 74.2 | 0.4 | 8:x on both |
| 82 | 48.5 | 0.5 | 80.6 | 1.2 | |
| 83 | 44.3 | -0.9 | 79.8 | 1.1 | |
| 84 | 49.4 | 0.8 | 75.0 | 0.5 | |
| 85 | 46.7 | -0.1 | 76.6 | 0.7 | |
| 86 | 52.7 | 1.9 | 72.6 | 0.2 | |
| 87 | 45.2 | -0.6 | 55.8 | -1.9 | |
| 88 | 48.8 | 0.6 | 79.0 | 1 | |
| 89 | 46.4 | -0.2 | 70.2 | -0.1 | |
| 90 | 49.7 | 0.9 | 82.2 | 1.4 |
Finally, here is the third set of Levey-Jennings charts for the values 60 through 90:
Remember the caveats:
So, at the end of this, can you answer the following questions?
[Did you miss something? Do you still feel like you're missing something? Please don't feel afraid to ask for more explanations.]