QC Selection Grid, Square 3 - Westgard QC

Tools, Technologies and Training for Healthcare Laboratories

QC Selection Grid, Square 3

MULTIRULE CHOICE:

Mqcsg3

You have many problems to detect, but fortunately medically important errors are large in size and may be detectable. It is recommended that you increase the number of rules to four and apply the 41s rule across two runs (R=2) as a rejection rule. The power curves below show the rejection characteristics within one run (R=1) and across two runs (R=2):


Mpf3

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curves. Where the curves and critical-error meet indicates how much error detection each control rule will provide.

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input