Quality - Westgard QC - Blog - Page 7 - Results from #120

Tools, Technologies and Training for Healthcare Laboratories

Q & A: Is this a 4:1s violation?

Posted by Sten Westgard, MS

We got the following question emailed to the website recently:

Greetings, I was [wondering] if the following case represents [violation] of 4 1s rule?

Control (A) point 1= EXCEEDS 2 SD /-2 SD lines >> warning

Control (A) point 2= WITHIN 2 SD/-2 SD lines >> accepted

Control (A) point 3= EXCEEDS 2 SD /-2 SD lines >> warning

Control (A) point 4= WITHIN 2 SD/-2 SD lines >> rejection

all the point[s] on the same side of the mean.

What do you think? The answer, after the jump

-----
Continue reading
  1424 Hits

Thank you, Jiao Tong University

Posted by Sten Westgard, MS

Continue reading
  421 Hits

Q & A: The basis of bias%?

Posted by Sten Westgard, MS

Continue reading
  632 Hits

Theranos Bleeds out...

Posted by Sten Westgard, MS

Continue reading
  396 Hits

Thank you, Hanoi

Posted by Sten Westgard, MS

Continue reading
  360 Hits

Q and A: Is the 10:x rule still compulsory?

A recent question came to me via social media: "Hello Sir, Good morning, How are you? I wanted to know is still 10x rule is coumpulsary?[stet.]  Even [when] we are using lab mean and lab SD. Please share the details.Thanks and Regards, [anonymous]"10:x is not compulsory by any regulation that I am aware of. Many regulations require you to...