Are there new (Erna Lenters-)"Westra Rules" for HbA1c? - Blog - Westgard QC

Tools, Technologies and Training for Healthcare Laboratories

Are there new (Erna Lenters-)"Westra Rules" for HbA1c?

Posted by Sten Westgard, MS

Erna Lenters-Westra and Emma English have a new paper in CCLM:

Evaluating new HbA1c methods for adoption by the IFCC and NGSP reference networks using international quality targets, Erna Lenters-Westra, Emma English,  2017 Aug 28;55(9):1426-1434. doi: 10.1515/cclm-2017-0109

They continue the standard set by the IFCC Task Force on Implementation of HbA1c standardization (TF-HbA1c) that advocates the Sigma-metric approach as the model for setting and evaluating quality targets for HbA1c. "The TF-HbA1c has set default risk levels of 2 [Sigma] for routine laboratories and 4 [Sigma] for laboratories performing clinical trials."

Further, the paper established its own allowable total error for evaluation:

"The total allowable error (TAE) for HBA1c has been set by the TF-HbA1c as a default of 5 mmol/mol (0.45% DCCT) at an HbA1c level of 50 mmol/mol (6.7% DCCT), which corresponds with a relative TAE of 10%... in SI units [6.9% DCCT units...]"

In conversation with Dr. Lenters-Westra, she has outlined a fleshed-out set of analytical performance acceptability specifications:

  • CV in CLSI EP-5 should be at 48 mmol/mol and 75 mmol/mol < or equal to 3% in SI units or < or equal to 2% in DCCT units
  • Bias in CLSI EP-9 at 48 mmol/mol and 75 mmol/mol compared with the mean of at least 3 certified secondary reference measurement procedures should < or equal to 2 mmol/mol or < or equal to 0.2% DCCT units.

Perhaps we should refer to these as (Lenters-)"Westra Rules" for HbA1c?

Want to hear more about this? We encourage you to attend the Quality in the Spotlight Conference in 2018, where Dr. Lenters-Westra will lecture extensively on her work (as well as receive the Westgard Award).

-----

×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Multisite Analysis of AST performance and goals
San Diego Smiles
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Thursday, 22 October 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input