You've read about this new Quality Planning process. But you're skeptical. Want to see proof? How about some real-world examples, showing bench-level quality planning in seconds. Well, this lesson provides an overview of needs, approach, and methodology for quality planning in healthcare laboratories. This overview should be useful as a preview, review, or quick refresher of the quality-planning process and the training materials available to support your applications.
Note: This material is covered in the Basic Planning for Quality manual. Updated coverage of these topics can be found in Assuring the Right Quality Right, as well as the Management and Design of Analytical Quality Systems online course.This lesson provides an overview of needs, approach, and methodology for quality planning in healthcare laboratories. This overview should be useful as a preview, review, or quick refresher of the quality-planning process and the training materials available to support your applications. Because we all learn best from tests and methods that are of interest to ourselves, the remaining lessons present a variety of example applications for routine chemistry, toxicology, hematology, endocrinology, and immunology. Use these applications to master the steps of the quality-planning process and become proficient in performing the process. Proficient means you should be able to select a QC procedure and identify the TQC strategy in one minute or less. Even the one-minute manager can perform quality planning with support of the tools and technology available today!
A series of fifteen lessons are available to support your understanding and application of quality planning for laboratory tests. A manual quality-planning process is available as part of these training materials and can be readily implemented in any healthcare laboratory. Mini-courses, short courses, and workshops may utilize only a subset of these lessons, so you may find it useful to know about all the lessons that are available.
The first three lessons in this series [1-3] focus on the WHY or need for quality planning, the next two [4-5] describe WHAT should be done to plan the quality of laboratory tests, and the following three [6-8] provide the HOW or practical methodology for performing quality planning quickly and easily. This lesson summarizes important points about the quality-planning process and provides additional discussion about WHEN to do quality planning and WHO should be involved and responsible. The remaining lessons provide example applications for different areas of the laboratory.
1. The laboratory must provide a final and independent assurance of quality for the test results it produces. This need for an independent quality management process is described in A Wake-up Call for Laboratory Quality Management [1], which discusses the limitations of current laboratory management practices that bundle many of the essential quality management activities into the services provided by manufacturers of analytical systems. A current FDA action against a major manufacturer exposes several fallacies in current management thinking:
2. The principles of Total Quality Management (TQM) point out the importance of defining quality standards and establishing a quality planning process that builds the desired quality into daily production processes. Assuring Quality through Total Quality Management [2] describes a quality management framework that is constructed from components - Quality Laboratory Processes, Quality Control, Quality Assessment, Quality Improvement, and Quality Planning - and assembled in a loop or cycle that provides continuous improvement of quality (as shown in the accompanying figure). This quality management framework should be centered on, or guided by, quality goals or standards. In most laboratories, implementation places a priority on the following:
3. Current regulatory and professional practice guidelines identify the need for quality planning and provide guidance for implementing a quality-planning process. In Complying with Regulations, Standards, and Practice Guidelines [3], the TQM framework is compared with JCAHO IOP guidelines (Joint Commision for Accreditation of Healthcare Organizations, Improving Organizational Performance) CLIA regulatory rules (Clinical Laboratory Improvement Amendments), and NCCLS QC practice guidelines (National Committee for Clinical Laboratory Standards), which reveals that:
4. A detailed quality-planning process can be developed by building on the NCCLS QC planning guidelines and introducing a quality-planning tool that provides information about the performance of different QC procedures. Devising a Practical Process [4] describes the steps of a quality-planning process that utilizing a graphical quality-planning tool - the chart of operating specifications, or OPSpecs chart. As shown in the accompanying figure, the two major areas of application are:
5. The first and most important step in the planning process is to define the quality required for the test. Unfortunately, this is not a trivial pursuit! Defining Quality Requirements [5] describes the difficulties and the solution:
6. A practical tool for translating quality requirements into operating specifications is needed to make quality planning quick and easy to perform. Adopting the OPSpecs Chart as Your Planning Tool [6] provides the basic details about a graphical tool called the chart of operating specifications (or OPSpecs chart), including:
7. Cost-effective operation of testing processes depends on individualizing the QC procedures for each test and method in the laboratory. QC is a technical management strategy and statistical QC is one component of the broad or Total QC strategy. Formulating a Total Quality Control Strategy [7] describes how to establish the appropriate balance between statistical and non-statistical components (preventive maintenance, instrument function checks, performance validation tests, patient data QC, in-service training) on the basis of the error detection capability of the statistical QC procedure for the application of interest. Three general strategies are pictured here:
8. Normalized OPSpecs charts provide a practical tool for doing quality planning by hand. Implementing a Manual Process using Normalized OPSpecs Charts [8] describes how OPSpecs charts can be scaled to be used with any quality requirement. As shown here, the y-axis is scaled from 0 to 100 and the x-axis is scaled from 0 to 50. The operating point is then calculated as a percent of the quality requirement and plotted on the normalized chart. By using this approach, only six charts are needed to select a QC procedure for any test. Everything you need to get started is available from the Internet:
9. The steps of the manual planning process using normalized OPSpecs charts are summarized in the accompanying diagram.
In a healthcare laboratory, quality planning should be an ongoing activity. It's never too late or too early to apply the quality planning process. You need to begin right away! Given that the quality-planning process can be fast and easy to perform - one minute or less, the biggest demand on your time will be the time needed to define the quality requirement for the test and the time needed to determine the imprecision and inaccuracy of your methods. Once you become proficient at using the quality-planning process, you can apply the process to accomplish the following:
In workshops and seminars, I often tell participants that I can teach fifth and six grade students to do quality planning and it would take less time than teaching adults. The point of that statement is to emphasize that we have a well-defined quality-planning process and it's easy to do if you follow the directions. If given a quality requirement, the OPSpecs charts for that requirement, and the figures for the method bias and CV, any fifth or sixth grader could plot the operating point and identify the line or lines above the point. The mechanics of using an OPSpecs chart are simple.
The things that are difficult in this process are (a) defining the quality requirement for the test and (b) determining the precision and accuracy of the method. Those steps require skilled and experienced laboratory scientists. In principle, the medical director of the laboratory has the responsibility for defining the quality requirements for the tests. In practice, that activity may be delegated to managers, supervisors, lead analysts, or quality specialists. In cases where CLIA has defined an allowable total error, that requirement can be used as the starting point. Anyone familiar with the CLIA regulations could look up the requirements.
Reliable estimates of method imprecision and inaccuracy depend on proper statistical analysis of sets of experimental data, initially from method validation studies and later on from routine QC, monthly peer-review comparisons, and periodic proficiency testing surveys. Here's where training and experience with method validation studies and routine QC are important. Some statistical skills are needed - at least the knowledge of which experiments and statistics provide reliable estimates of different types of analytical errors. In many laboratories, these skills reside with QA or QC specialists, R&D technologists, senior analysts, or experienced managers and supervisors.
Cholesterol is my favorite test! I use cholesterol as an example over and over and over because it illustrates many of the problems and difficulties you can expect to encounter with other tests. As a clinical chemist, I've followed the evolution of cholesterol methodology and quality standards for over thirty years. In many ways, cholesterol provides a model system that illustrates the importance of analytical quality management in making test results commutable, i.e., interchangable from method to method and lab to lab [9].
Commutability should be our goal for all laboratory tests, not just cholesterol. This will require that each laboratory carefully plan it's testing processes to assure the quality required by national standards and guidelines. The cholesterol story shows the importance of having methods that are highly specific, excellent analytical performance (imprecision and inaccuracy), traceability of standards and calibrators, comparability between field methods and a national reference method, and statistical QC for assuring the quality of routine test results.
Few tests and methods are as well characterized and well studied as cholesterol. Therefore, we can expect there will be difficulties in assuring the quality of many of the other tests being performed in our laboratories. Even so, we will gain many useful insights into our quality management problems if we apply the quality-planning process to all our tests. To help get started with tests you're familiar with, the remaining lessons provide example applications in the following areas: