4:1s, when is it violated under different types of Westgard Rules? - Blog - Westgard QC

Tools, Technologies and Training for Healthcare Laboratories

4:1s, when is it violated under different types of Westgard Rules?

The Westgard Rules, in the classic 1981 multirule paper, require a 1:2s violation before other rejection rules are violated.  If you had4 values past 1s, but nothing past 2s, it wouldn't be checked. Under more "modern" Westgard Rules, with the 1:2s warning rule abolished, you don't need to wait until the 2s limit line is violated before interpreting the rule. Under Westgard Sigma Rules, it's further simplified: you don't need to interpret the 4:1s unless the method is 4 Sigma or lower. In other words, for a method of 5 or 6 Sigma quality, neither of the situations below would be considered a rejection-worthy event.

The 4:1s rule for the "classic" Westgard multirule
A 4:1s violation under a modern Westgard multirule
×
Stay Informed

When you subscribe to the blog, we will send you an e-mail when there are new updates on the site so you wouldn't miss them.

Question and Answer: How to use the 7:T rule?
Repeating the control: just more info, or just rol...
 

Comments

No comments made yet. Be the first to submit a comment
Already Registered? Login Here
Thursday, 22 October 2026

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input