In 2016, INVITRO Laboratories became the first (and largest) private laboratory in Russia to achieve the Verification of Performance. (Sigma VP). More than 40 analytes of two of their chemistry instruments were verified. In 2017, they re-verified this performance. In 2018, they expanded their verification to 4 other labs across Russia, encompassing 129 assays.
In 2016, Westgard QC was pleased to verify the performance of more than 45 assays at INVITRO, the largest private medical company in Russia.
INVITRO has 9 modern labs and owns the biggest network in Eastern Europe with over 900 medical offices in Russia, Ukraine, Kazakhstan, and Belarus. It was the first laboratory in Russia to obtain ISO 15189:2012 certification and ISO 9001:2011 (ISO 9001:2008). It participates in international external quality assurance programs that include FSVOK (Russia), EQAS - USA, and RIQAS - UK.
This pilot verification assessed the performance of two of their main core laboratory biochemistry instruments, the c8000 mid-volume chemistry analyzer and the c16000 high volume chemistry analyzer
In 2018, INVITRO expanded their pilot programs to 4 other branches in their network, further expanding to 5 branches in 2019.
|
Chemistry Assays Moscow |
C1601772 Verified |
C1601770 Verified |
|
Albumin |
Six Sigma |
Six Sigma |
|
Alkaline Phosphatase |
Six Sigma |
Six Sigma |
|
ALT |
Six Sigma |
Six Sigma |
|
Amylase |
Six Sigma |
Six Sigma |
|
AST |
Six Sigma |
Five Sigma |
|
Bilirubin, Direct |
Six Sigma |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
Six Sigma |
|
Calcium |
Six Sigma |
Six Sigma |
|
Chloride |
Five Sigma |
Four Sigma |
|
Cholesterol |
Six Sigma |
Five Sigma |
|
Creatinine Kinase |
Six Sigma |
Six Sigma |
|
Creatinine |
Six Sigma |
Six Sigma |
|
GGT |
Four Sigma |
Six Sigma |
|
Glucose |
Four Sigma |
Five Sigma |
|
HDL |
Five Sigma |
Six Sigma |
|
Iron |
Five Sigma |
Four Sigma |
|
LDH |
Four Sigma |
Four Sigma |
|
Magnesium |
Five Sigma |
Six Sigma |
|
Phosphorous |
Six Sigma |
Six Sigma |
|
Potassium |
Six Sigma |
Six Sigma |
|
Protein, Total |
Six Sigma |
Six Sigma |
|
Sodium |
Four Sigma |
- |
|
Triglycerides |
Six Sigma |
Six Sigma |
|
Uric Acid |
Six Sigma |
Six Sigma |
|
Urine Assays Moscow |
C1601770 Verified |
|
Amylase, U |
Six Sigma |
|
Calcium, U |
Six Sigma |
|
Glucose, U |
Six Sigma |
|
Magnesium, U |
Six Sigma |
|
Phosphorous, U |
Six Sigma |
|
Potassium, U |
Six Sigma |
|
Protein, U |
Six Sigma |
|
Sodium, U |
Six Sigma |
|
Urea N, U |
Six Sigma |
|
Uric, U |
Six Sigma |
|
Immunoassays Moscow |
AI23004 Verified |
AI23005 Verified |
|
Cortisol |
Four Sigma |
- |
|
Estradiol |
Five Sigma |
- |
|
FSH |
Five Sigma |
Five Sigma |
|
LH |
Six Sigma |
- |
|
Progesterone |
Six Sigma |
- |
|
Prolactin |
Six Sigma |
- |
|
Testosterone |
Six Sigma |
Six Sigma |
|
Chemistry Assays Siberia |
C804275 Verified |
|
Albumin |
Six Sigma |
|
Alkaline Phosphatase |
Six Sigma |
|
ALT |
Six Sigma |
|
Amylase |
Six Sigma |
|
AST |
Five Sigma |
|
Bilirubin, Direct |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
|
Calcium |
Six Sigma |
|
Chloride |
Five Sigma |
|
Cholesterol |
Six Sigma |
|
Creatinine Kinase |
Six Sigma |
|
Creatinine |
Six Sigma |
|
GGT |
Six Sigma |
|
Glucose |
Six Sigma |
|
HDL |
Six Sigma |
|
Iron |
Four Sigma |
|
LDL |
Six Sigma |
|
Lipase |
Six Sigma |
|
Magnesium |
Five Sigma |
|
Phosphorous |
Six Sigma |
|
Potassium |
Six Sigma |
|
Protein, Total |
Six Sigma |
|
Transferin |
Six Sigma |
|
Triglycerides |
Six Sigma |
|
UIBC |
Five Sigma |
|
Urea Nitrogen |
Five Sigma |
|
Uric Acid |
Six Sigma |
|
Immunoassays Siberia |
ISR61139 Verified |
ISR61143 Verified |
AI22405 Verified |
|
Anti-TG |
- |
- |
Five Sigma |
|
Anti-TPO |
- |
- |
Six Sigma |
|
AFP |
- |
Six Sigma |
- |
|
Estradiol |
Six Sigma |
- |
- |
|
Folate |
Four Sigma |
- |
- |
|
LH |
Four Sigma |
- |
- |
|
Progesterone |
Six Sigma |
- |
- |
|
Prolactin |
- |
Six Sigma |
- |
|
iPTH |
- |
- |
Five Sigma |
|
T3, Total |
- |
- |
Four Sigma |
|
T4, Free |
- |
- |
Five Sigma |
|
T4, Total |
- |
- |
Five Sigma |
|
Testosterone |
- |
Five Sigma |
- |
|
Chemistry Assays Ural |
804573 Verified |
C804891 Verified |
|
Albumin |
Six Sigma |
Six Sigma |
|
Alkaline Phosphatase |
Six Sigma |
Six Sigma |
|
ALT |
Six Sigma |
Six Sigma |
|
Amylase |
Six Sigma |
Six Sigma |
|
AST |
Four Sigma |
Four Sigma |
|
Bilirubin, Direct |
Six Sigma |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
Five Sigma |
|
C3 |
Six Sigma |
- |
|
C4 |
Six Sigma |
- |
|
C-Reactive Protein |
Six Sigma |
- |
|
Calcium |
Five Sigma |
Five Sigma |
|
Chloride |
Four Sigma |
Five Sigma |
|
Cholesterol |
Four Sigma |
Six Sigma |
|
CO2 |
- |
Four Sigma |
|
Creatinine Kinase |
Six Sigma |
- |
|
Creatinine |
Six Sigma |
Six Sigma |
|
GGT |
Six Sigma |
Six Sigma |
|
Glucose |
Four Sigma |
Four Sigma |
|
HDL |
Six Sigma |
Six Sigma |
|
IgA |
Six Sigma |
- |
|
IgG |
Six Sigma |
- |
|
IgM |
Six Sigma |
- |
|
Iron |
Six Sigma |
Five Sigma |
|
Lactate |
Four Sigma |
- |
|
LDL |
Six Sigma |
- |
|
LDH |
- |
Five Sigma |
|
Lipase |
Five Sigma |
- |
|
Magnesium |
- |
Five Sigma |
|
Phosphorous |
Six Sigma |
Six Sigma |
|
Potassium |
Six Sigma |
Six Sigma |
|
Protein, Total |
Six Sigma |
Six Sigma |
|
Sodium |
- |
Four Sigma |
|
Triglycerides |
Six Sigma |
Six Sigma |
|
UIBC |
Four Sigma |
- |
|
Uric Acid |
Four Sigma |
Five Sigma |
|
Urine Assays Ural |
C804891 Verified |
|
Amylase, U |
Six Sigma |
|
Calcium, U |
Six Sigma |
|
Creatinine, U |
Six Sigma |
|
Glucose, U |
Six Sigma |
|
Magnesium, U |
Six Sigma |
|
Phosphorous, U |
Six Sigma |
|
Protein, U |
Six Sigma |
|
Urea N, U |
Six Sigma |
|
Uric, U |
Six Sigma |
|
Immunoassays Ural |
AI05057 Verified |
AI21777 Verified |
|
AFP |
- |
Four Sigma |
|
FSH |
Five Sigma |
- |
|
LH |
Five Sigma |
- |
|
Progesterone |
Five Sigma |
- |
|
T4, Free |
Five Sigma |
- |
|
T4, Total |
Five Sigma |
- |
|
Prolactin |
- |
Six Sigma |
|
PSA |
- |
Five Sigma |
|
Testosterone |
Six Sigma |
- |
|
TSH |
Six Sigma |
- |
|
Tumor Markers assays Ural |
AI21777 Verified |
|
CA 125 II |
Four Sigma |
|
CA 15-3 |
Four Sigma |
|
CA 19-9 XR |
Four Sigma |
|
Chemistry Assays Kazakhstan |
C804691 Verified |
|
Albumin |
Six Sigma |
|
Alkaline Phosphatase |
Five Sigma |
|
ALT |
Six Sigma |
|
Amylase |
Six Sigma |
|
AST |
Five Sigma |
|
Bilirubin, Direct |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
|
Calcium |
Four Sigma |
|
Chloride |
Four Sigma |
|
Creatinine |
Six Sigma |
|
GGT |
Six Sigma |
|
Glucose |
Six Sigma |
|
HDL |
Five Sigma |
|
Iron |
Six Sigma |
|
LDL |
Six Sigma |
|
Magnesium |
Five Sigma |
|
Potassium |
Six Sigma |
|
Protein, Total |
Six Sigma |
|
Triglycerides |
Six Sigma |
|
Urea Nitrogen |
Six Sigma |
|
Uric Acid |
Six Sigma |
|
Immunoassays Kazakhstan |
I63209 Verified |
I63210 Verified |
|
Anti-TG |
- |
Six Sigma |
|
Anti-TPO |
- |
Six Sigma |
|
Estradiol |
Five Sigma |
- |
|
Ferritin |
Six Sigma |
- |
|
Folate |
Six Sigma |
- |
|
FSH |
Five Sigma |
- |
|
Insulin |
Six Sigma |
- |
|
Progesterone |
Four Sigma |
- |
|
Prolactin |
Six Sigma |
- |
|
T3, Free |
- |
Four Sigma |
|
T3, Total |
- |
Four Sigma |
|
T4, Free |
- |
Six Sigma |
|
T4, Total |
- |
Four Sigma |
|
Testosterone |
Six Sigma |
- |
|
TSH |
- |
Six Sigma |
|
Tumor Markers assays Kazakhstan |
I63209 Verified |
|
CA 125 II |
Six Sigma |
|
CA 15-3 |
Four Sigma |
|
CA 19-9 XR |
Five Sigma |
|
Chemistry Assays Belarus |
C804526 Verified |
C804627 Verified |
|
Albumin |
- |
Six Sigma |
|
Alkaline Phosphatase |
Six Sigma |
Six Sigma |
|
ALT |
Six Sigma |
Six Sigma |
|
Amylase |
- |
Six Sigma |
|
AST |
Six Sigma |
Five Sigma |
|
Bilirubin, Direct |
Six Sigma |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
Six Sigma |
|
C-Reactive Protein |
Six Sigma |
Six Sigma |
|
Calcium |
- |
Four Sigma |
|
Chloride |
Four Sigma |
- |
|
Cholesterol |
Six Sigma |
Six Sigma |
|
Creatinine Kinase |
- |
Six Sigma |
|
Creatinine |
Six Sigma |
Six Sigma |
|
GGT |
Six Sigma |
Six Sigma |
|
Glucose |
Four Sigma |
Four Sigma |
|
HDL |
Four Sigma |
Six Sigma |
|
Iron |
- |
Five Sigma |
|
LDL |
Six Sigma |
- |
|
LDH |
- |
Four Sigma |
|
Lipase |
- |
Six Sigma |
|
Magnesium |
- |
Six Sigma |
|
Phosphorous |
- |
Six Sigma |
|
Potassium |
Six Sigma |
- |
|
Protein, Total |
Five Sigma |
Four Sigma |
|
Transferin |
- |
Six Sigma |
|
Triglycerides |
Six Sigma |
Six Sigma |
|
Uric Acid |
- |
Six Sigma |
|
Urine Assays Belarus |
C804526 Verified |
C804627 Verified |
|
Creatinine, U |
- |
Six Sigma |
|
Glucose, U |
- |
Six Sigma |
|
Protein, U |
Six Sigma |
- |
|
Uric, U |
- |
Six Sigma |
|
Immunoassays Belarus |
I62718 Verified |
I62753 Verified |
|
Cortisol |
Five Sigma |
- |
|
Estradiol |
Six Sigma |
- |
|
Ferritin |
- |
Four Sigma |
|
Folate |
- |
Six Sigma |
|
Insulin |
- |
Six Sigma |
|
LH |
Four Sigma |
- |
|
Progesterone |
Four Sigma |
- |
|
Prolactin |
Four Sigma |
- |
|
T3, Free |
- |
Four Sigma |
|
T3, Total |
- |
Four Sigma |
|
T4, Free |
- |
Six Sigma |
|
T4, Total |
- |
Six Sigma |
|
Testosterone |
Six Sigma |
- |
|
Vitamin B12 |
- |
Four Sigma |
|
Chemistry Assays St. Petersburg |
C804721 Verified |
|
Albumin |
Six Sigma |
|
Alkaline Phosphatase |
Six Sigma |
|
ALT |
Six Sigma |
|
Amylase |
Six Sigma |
|
AST |
Six Sigma |
|
Bilirubin, Direct |
Six Sigma |
|
Bilirubin, Total |
Six Sigma |
|
C-Reactive Protein |
Six Sigma |
|
Calcium |
Six Sigma |
|
Chloride |
Five Sigma |
|
Cholesterol |
Six Sigma |
|
Creatinine Kinase |
Six Sigma |
|
Creatinine |
Six Sigma |
|
GGT |
Six Sigma |
|
Glucose |
Five Sigma |
|
HDL |
Six Sigma |
|
IgA |
Six Sigma |
|
IgG |
Six Sigma |
|
IgM |
Six Sigma |
|
Iron |
Four Sigma |
|
LDL |
Four Sigma |
|
LDH |
Four Sigma |
|
Lipase |
Five Sigma |
|
Magnesium |
Five Sigma |
|
Phosphorous |
Six Sigma |
|
Potassium |
Six Sigma |
|
Protein, Total |
Six Sigma |
|
Sodium |
Four Sigma |
|
Transferin |
Six Sigma |
|
Triglycerides |
Five Sigma |
|
UIBC |
Six Sigma |
|
Urea Nitrogen |
Four Sigma |
|
Uric Acid |
Six Sigma |
|
Urine Assays St. Petersburg |
C804721 Verified |
|
Creatinine, U |
Six Sigma |
|
Glucose, U |
Six Sigma |
|
Protein, U |
Six Sigma |
|
Uric, U |
Six Sigma |
|
Immunoassays St. Petersburg |
AI04567 Verified |
ISR63357 Verified |
ISR56113 Verified |
|
Anti-TG |
Four Sigma |
- |
- |
|
Anti-TPO |
Six Sigma |
- |
- |
|
AFP |
- |
- |
Five Sigma |
|
Cortisol |
- |
- |
Five Sigma |
|
Estradiol |
- |
Six Sigma |
- |
|
Ferritin |
- |
- |
Four Sigma |
|
Folate |
- |
- |
Four Sigma |
|
FSH |
- |
Five Sigma |
- |
|
Insulin |
- |
- |
Six Sigma |
|
LH |
- |
Five Sigma |
- |
|
Progesterone |
- |
Six Sigma |
- |
|
Prolactin |
- |
Six Sigma |
- |
|
iPTH |
Five Sigma |
- |
- |
|
T3, Free |
Five Sigma |
- |
- |
|
T3, Total |
Four Sigma |
- |
- |
|
T4, Free |
Six Sigma |
- |
- |
|
T4, Total |
Five Sigma |
- |
- |
|
Testosterone |
- |
Six Sigma |
- |
|
TSH |
Six Sigma |
- |
- |
|
Vitamin B12 |
- |
- |
Five Sigma |
|
Tumor Markers assays St. Petersburg |
AI04567 Verified |
|
CA 125 II |
Four Sigma |
|
CA 15-3 |
Four Sigma |
|
CA 19-9 XR |
Four Sigma |
|
Chemistry Assays Samara |
C802571 Verified |
|
Alkaline Phosphatase |
Five Sigma |
|
ALT |
Six Sigma |
|
Amylase |
Five Sigma |
|
AST |
Four Sigma |
|
Bilirubin, Direct |
Six Sigma |
|
Bilirubin, Total |
Five Sigma |
|
C-Reactive Protein |
Six Sigma |
|
Cholesterol |
Four Sigma |
|
Creatinine |
Six Sigma |
|
HDL |
Six Sigma |
|
Iron |
Five Sigma |
|
LDL |
Four Sigma |
|
Phosphorous |
Six Sigma |
|
Transferin |
Six Sigma |
|
Triglycerides |
Four Sigma |
|
Immunoassays Samara |
I60295 Verified |
AI04150 Verified |
|
Anti-TPO |
- |
Six Sigma |
|
Ferritin |
- |
Four Sigma |
|
FSH |
Four Sigma |
- |
|
Insulin |
- |
Six Sigma |
|
Progesterone |
Four Sigma |
- |
|
Prolactin |
Six Sigma |
- |
|
Testosterone |
Six Sigma |
- |
|
TSH |
- |
Six Sigma |
|
Vitamin B12 |
- |
- |



































Congratulations to INVITRO on their accomplishments.
Sten Westgard, MS, pictured here in 2016 with INVITRO founder and Chairman of the Board, Alexander Ostrovskiy.

Pictured here on the right is Elena Kondrasheva, Laboratory Director.