Tools, Technologies and Training for Healthcare Laboratories

Medical Decision Levels

These tables of medical decision levels provide possible critical decision levels - where you can assess performance (CV, bias) and determine the Sigma-metrics and appropriate QC procedures.

These figures are quoted from Statland BE. Clinical Decision Levels for Laboratory Tests, Second Edition [Oradell NJ;Medical Economics Books, 1987.] with permission.

MEDICAL DECISION LEVELS

These figures are quoted from Statland BE. Clinical Decision Levels for Laboratory Tests, Second Edition [Oradell NJ;Medical Economics Books, 1987.] with permission.

Test Units Reference
Interval
Decision Levels
ELECTROLYTES 1 2 3 4 5
Calcium mg/dL 9.0-10.6 7.0 11.0 13.5
Chloride mmol/L 98-109 90 112
CO2 Content mmol/L 23-30 6.0 20 33
Magnesium mEq/L 1.2-2.4 1.2 2.0 5.0
mmol/L 0.6-1.2 0.6 1.0 2.5
Phosphorus mg/dL 2.5-5.0 1.5 2.5 5.0
Potassium mmol/L 3.7-5.1 3.0 5.8 7.5
Sodium mmol/L 138-146 115 135 150

Back to top


Test Units Reference
Interval
Decision Levels
METABOLITES 1 2 3 4 5
Bilirubin mg/dL 0.1-1.2 1.4 2.5 20
Cholesterol mg/dL 150-275 90 240 260 350
Creatinine mg/dL 0.7-1.5 0.6 1.6 6.0
Glucose mg/dL 60-95 45 120 180
Iron ug/dL 50-165 50 220 400
Triglycerides mg/dL 20-180 40 150 400
Urea Nitrogen (BUN) mg/dL 8-26 6 26 50
Uric acid mg/dL 2.5-7.0 2.0 8.0 10.7

Back to top

Test Units Reference
Interval
Decision Levels
PROTEINS and ENZYMES 1 2 3 4 5
Alanine aminotransferase U/L 5-40 20 60 300
Albumin g/dL 3.5-5.0 2.0 3.5 5.2
Alkaline phosphatase U/L 35-120 50 150 400
Amylase Somogyi U 60-180 50 120 200
Aspartate aminotransferase U/L 8-40 20 60 300
Carcinoembryonic antigen ng/mL <2.5 2.5 10 20
Creatine kinase U/L 10-180 100 240 1800
Glutamyl transferase U/L 5-40 20 50 150
Lactate dehydrogenase U/L 60-220 150 300 500
Total protein g/dL 6.0-8.0 4.5 6.0 8.0

Back to top

Test Units Reference
Interval
Decision Levels
HORMONES 1 2 3 4 5
Cortisol in plasma ug/dL 7-20 @8am 5 10 18 25
Cortisol (Free) in urine ug/24 hrs 20-90 50 100
Follicle-stimulating hormone mIU/mL 2-15 Adult M 1 35 60
Follicular phase, Luteal phase 3-15 Adult F
17-Hydroxycorticosteroids mg/24 hrs 3-10 3 15 20
17-Ketosteroids in urine mg/24 hrs 8-20 Adult M 20 25 40
5-15 Adult F
Luteinizing hormone mIUL/mL 5-25 Adult M 1 50 100
Follicular phase 5-30 Adult F
Metanephrine in urine mg/24 hrs <1.3 1.3 2.5
Prolactin ng/mL 1-20 M 30 100 300
1-25 F
Thyroxine ug/dL 5.5-12.5 5 7 10 14
Vanillymandelic acid in urine mg/24 hrs <6.8 6.8 13

Back to top

Test Units Reference
Interval
Decision Levels
HEMATOLOGY RELATED TESTS 1 2 3 4 5
Antithrombin-III % of normal 80-120 50 75
Bleeding Time min 2.3-9.2 10 15
Fibrinogen in plasma mg/dL 200-400 30 100 500
Folate in serum ng/mL 2-15 1.5 4.0
Hematocrit L/L 0.43-0.51 M 0.14 0.33 0.56 0.70
0.38-0.46 F
Hemoglobin g/dL 14-17.8 M 4.5 10.5 17 23
12-15.6 F
Mean corpuscular volume fL (cu u) 84-96 80 100
Partial thromboplastin time sec 30 35 45 90
Plasminogen % 80-120 50 75 135
Platelet count K/uL 150-400 10 50 100 600 1000
Prothrombin time sec 11.5 14 16 30
Vitamin B12 pg/mL 200-900 170 250 1200
White blood cell count K/uL 4-11 0.5 3 12 30

Back to top

Test Units Theraputic
Range
Decision Levels
DRUGS 1 2 3 4 5
Amikacin ug/mL 20-30 (peak) 10 20 32
1-8 (trough)
Carbamazepine ug/mL 4-12 4 9 14
Digoxin ug/mL 0.9-2.0 1.0 1.6 2.5
Disopyramide ug/mL 2.5-5.0 2.5 4.5 7.0
Ethosuximide ug/mL 50-100 50 120
Gentamicin ug/mL 6-10 (peak) 0.5 2 6 12
0.5-1.5 (trough)
Kanamycin ug/mL 20-30 (peak) 10 20 32
1-8 (trough)
Lidocaine ug/mL 1.5-5.0 1.5 5.0 7.0
Lithium mmol/L 0.5-1.2 0.4 0.8 1.5
Phenobarbital ug/mL 15-40 15 30 60 120
Phenytoin ug/mL 10-20 10 20 50
Primidone ug/mL 5-12 5 9 12
Procainamide ug/mL 4-10 4 12 16
Quinidine ug/mL 2-5 2 5 7
Salicylates ug/mL 150-300 150 300 500
Theophylline ug/mL 10-20 asthma 10 20 35 60
5-10 neonatal apnea
Tobramycin ug/mL 6-10 (peak) 0.5 2 6 12
0.5-1.5 (trough)
Valproic acid ug/mL 50-100 50 90 120

Back to top

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input