RCPA Allowable Limits of Performance for Biochemistry

The RCPA (Royal College of Pathologists of Australasia) have developed a thorough set of specifications for allowable error for biochemistry. For those laboratories not bound by the CLIA regulations, these can serve as a valuable resource. Many analytes that are non-regulated by CLIA can be found here, too.

Royal College of Pathologists of Australasia Analytical Quality Requirements


The tables below contain information on quality requirements set by the Royal College of Pathologists of Australasia and the Australasian Clincal Biochemist association Quality Assurance Program. These guidelines for acceptable performance can be used as Analytical Quality Requirements in the QC Design and Planning process.

[Please note: This is an unauthorized listing of RCPA allowable limits of performance (ALP). For official and up-to-date goals, consult the RCPA directly.]

2010 Update: In August, the AACB Scientific and Regulatory Affairs Committee updated the guidelines, tightening the allowable limits of performance for 21 analytes, as well as loosened the limits for 8 analytes. 

 [Further updates were made in 2012, 2013, 2014. These changes were belatedly updated here in 2020.]

You can skip ahead to specific areas of laboratory tests by clicking on the listed items:

Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Alcohol/Ammonia - updated 2012
Test or Analyte Lower Goal Upper Goal
Alcohol ± 2.0 mmol/L < 20.0 mmol/L ± 10% > 20.0 mmol/L
Ammonia ± 5 umol/L < 50 umol/L ± 20% > 50 umol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage.Some browsers may not display the plus/minus character (it shouldappear right here, ±. If it does not, your browser doesnot support this character).
Antibiotics - updated 2013
Test or Analyte Lower Goal Upper Goal
Amikacin ± 2.0 mg/L < 19.9 mg/L ± 10% > 19.9 mg/L
Gentamicin ± 0.2 mg/L < 2.0 mg/L ± 10% > 2.0 mg/L
Tobramycin ± 0.2 mg/L < 2.0 mg/L ± 10% > 2.0 mg/L
Vancomycin ± 2.0 mg/L < 20.3 mg/L ± 10% > 20.3 mg/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).
Biogenic Amines - updated 2012
Test or Analyte Lower Goal Upper Goal
Adrenaline ± 30 nmol/day < 100 nmol/day ± 30% > 100 nmol/day
Dopamine ± 0.20 umol/day < 2.0 umol/day ± 10% > 2.0 umol/day
5HIAA ± 8 umol/day < 40 umol/day ± 20% > 40 umol/day
HMMA ± 6 umol/day < 40 umol/day ± 15% > 40 umol/day
HVA ± 6 umol/day < 40 umol/day ± 15% > 40 umol/day
Metanephrine ± 0.2 umol/day < 1.0 umol/day ± 20% > 1.0 umol/day
Noradrenaline ± 75 nmol/day < 500 nmol/day ± 15% > 500 nmol/day
Normetanephrine ± 0.4 umol/day < 2.0 umol/day ± 20% > 2.0 umol/day
3 - Methoxytyramine ± 0.3 umol/day < 2.0 umol/day  ± 15% > 2.0 umol/day
Serotonin ± 0.2 umol/day < 1.0 umol/day ± 20% > 1.0 umol/day

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Blood Gas - updated 2012
Test or Analyte Lower Goal Upper Goal
Chloride
± 3.0 mmol/L
Glucose ± 1.0 mmol/day < 10.0 mmol/day ± 10% > 10.0 mmol/day
Ionised Calcium
± 0.05 mmol/L
Lactate ± 1.0 mmol/day < 10.0 mmol/day ± 10% > 10.0 mmol/day
pH
± 0.04
pCO2 ± 2.0 mm Hg < 25.0 mm Hg ± 8% > 25.0 mm Hg
pO2 ± 5 mm Hg < 100 mm Hg ± 5% > 100 mm Hg
Potassium
± 0.2 mmol/L
Sodium
± 3.0 mmol/L
Urea ± 1.0 < 10.0 mmol/L ± 10% > 10.0 mmol/L
Creatinine ± 10 < 100 mmol/L ± 10% > 100 mmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Co-oximetry
Test or Analyte Lower Goal Upper Goal
Hemoglobin concentration ± 3 g/L < 100 g/L ± 3% > 100 g/L
Fractional Oxyhemoglobin ± 3 < 75% ± 4% > 75%
Fractional Carboxyhemoglobin
± 2.0%
Fractional Methemoglobin ± 1.0 < 10% ± 10% > 100%

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

CSF
Test or Analyte Lower Goal Upper Goal
Albumin ± 0.02 g/L < 0.45 g/L ± 5% > 0.45 g/L
Glucose ± 0.2 mmol/L < 2.0 mmol/L ± 10% > 2.0 mmol/L
Immunoglobin G ± 0.02 g/L < 0.1 g/L ± 20% > 0.1 g/L
Lactate ± 0.3 mmol/L < 3.0 mmol/L ± 10% > 3.0 mmol/L
Total Protein ± 0.02 g/L < 0.45 g/L ± 5% > 0.45 g/L
Bilirubin Concentration ± 0.12 umol/L < 0.6 umol/L ± 20% > 0.6 umol/L
Xanthochromia - Bilirubin screen ± 0.002 AU < 0.007 AU ± 20% > 0.007 AU
Xanthochromia - Haemoglobin screen ± 0.02 AU < 0.1 AU ± 20% > 0.1 AU

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Endocrine - updated 2012
Test or Analyte Lower Goal Upper Goal
AFP ± 2 kIU/L < 17 kIU/L ± 12% > 17 kIU/L
Aldosterone ± 24 pmol/L < 160 pmol/L ± 15% > 160 pmol/L
Androstenedione ± 1.5 nmol/L < 10 nmol/L ± 15% > 10 nmol/L
CA125 ± 6 kU/L < 50 kU/L ± 12% > 50 kU/L
CEA ± 0.6 ug/L < 5.0 ug/L ± 12% > 5.0 ug/L
Cortisol ± 15 nmol/L < 100 nmol/L ± 15% > 100 nmol/L
11-Deoxycortisol (pre 2012) ± 30 nmol/L < 200 nmol/L ± 15% > 200 nmol/L
DHEA Sulphate ± 1.2 umol/L < 10 umol/L ± 12% > 10 umol/L
Ferritin ± 4.0 ug/L < 27.0 ug/L ± 15% > 27.0 ug/L
Folate

± 1.5 umol/L < 6.0 umol/L

± 25% > 6.0 umol/L
FSH ± 1.0 IU/L < 10.0 IU/L ± 10% > 10.0 IU/L
Gastrin (pre 2012) ± 25 pmol/L < 250 pmol/L ± 10% > 250 pmol/L
Growth Hormone ± 1 mU/L < 7 mU/L ± 15% > 7 mU/L
hCG ± 1 IU/L < 10.0 IU/L ± 10% > 10.0 IU/L
Homocysteine ± 1.5 umol/L < 15.0 umol/L ± 10% > 15.0 umol/L
17-Hydroxyprogesterone ± 2 nmol/L < 10 nmol/L ± 20% > 10 nmol/L
IgE (pre 2012) ± 4.0 IU/L < 20.0 IU/L ± 20% > 20.0 IU/L
Insulin ± 0.6 mU/L < 5.0 mU/L ± 12% > 5.0 mU/L
LH ± 1.50 IU/L < 10.0 IU/L ± 15% > 10.0 IU/L
Oestradiol ± 25 pmol/L < 100 pmol/L ± 25% > 100 pmol/L
Oestriol(unconjugated) ± 0.9 nmol/L < 6.0 nmol/L ± 15% > 6.0 nmol/L
PTH ± 1.0 pmol/L < 8.0 pmol/L ± 12% > 8.0 pmol/L
Progesterone ± 2 nmol/L < 10 nmol/L ± 15% > 10 nmol/L
Prolactin ± 40 mIU/L < 400 mIU/L ± 10% > 400 mIU/L
SHBG ± 6 nmol/L < 50 nmol/L ± 12% > 50 nmol/L
Testosterone ± 0.4 nmol/L < 2.7 nmol/L ± 15% > 2.7 nmol/L
TSH ± 0.1 mU/L < 0.5 mU/L ± 20% > 0.5 mU/L
Free T3 ± 0.7 pmol/L < 3.5 pmol/L ± 20% > 3.5 pmol/L
Free T4 ± 1.5 pmol/L < 12 pmol/L ± 12% > 12 pmol/L
Total T3 ± 0.2 nmol/L < 1.3 nmol/L ± 15% > 1.3 nmol/L
Total T4 ± 8 nmol/L < 50 nmol/L ± 15% > 50 nmol/L
Total Thyroxine ± 12 nmol/L < 120 nmol/L ± 10% > 120 nmol/L
Vitamin B12 ± 18 pmol/L < 120 pmol/L ± 15% > 120 pmol/L
Vitamin D3 ± 9 nmol/L < 60 nmol/L ± 15% > 60 nmol/L
Total PSA ± 0.4 ug/L < 5.0 ug/L ± 8% > 5.0 ug/L
Free PSA ± 0.2 ug/L < 1.4 ug/L ± 15% > 1.4 ug/L
ACE ± 1.0 U/L < 10.0 U/L ± 10% > 10.0 U/L
Red Cell Folate ± 80 nmol/L < 400 nmol/L ± 20% > 400 nmol/L
Plasma Renin Activity (pre 2012) ± 1.0 ng/mL/hr; < 4.0 ng/mL/hr ± 25% > 4.0 ng/mL/hr
Active Renin (pre 2012) ± 1.0 uU/mL < 4.0 uU/mL ± 25% > 4.0 uU/mL

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

General Serum Chemistry - updated 2012
Test or Analyte Lower Goal Upper Goal
Acid Phos(Total) (pre 2012) ± 1.5 U/L & ug/L < 10 U/L & ug/L ± 15% > 25 U/L & ug/L
Acid Phos(Pros) (pre 2012) ± 1.5 U/L & ug/L < 10 U/L & ug/L ± 15% > 25 U/L & ug/L
Albumin ± 2.0 g/L < 33.0 g/L ± 6% > 33 g/L
Alkaline Phosphatase ± 15 U/L < 125 U/L ± 12% > 125 U/L
ALT ± 5 U/L < 40 U/L ± 12% > 40 U/L
Amylase ± 10 U/L < 100 U/L ± 10% > 100 U/L
AST ± 5 U/L < 40 U/L ± 12% > 40 U/L
Bicarbonate ± 2.0 mmol/L < 20 mmol/L ± 10% > 20 mmol/L
Bilirubin - Total ± 3 umol/L < 25 umol/L ± 12% > 25 umol/L
Bilirubin Conjugated ± 3 umol/L < 15 umol/L ± 20% > 15 umol/L
Calcium ± 0.10 mmol/L < 2.5 mmol/L ± 4% > 2.5 mmol/L
Chloride ± 3.0 mmol/L < 100 mmol/L ± 3% > 100 mmol/L
Cholesterol ± 0.3 mmol/L < 5 mmol/L ± 6% > 5 mmol/L
Cholinesterase ± 500 U/L < 5000 U/L ± 10% > 5000 U/L
Creatine Kinase ± 15 U/L < 125 U/L ± 12% > 125 U/L
CK-MB ± 3 U/L & ug/L < 15 U/L & ug/L ± 20% > 15 U/L & ug/L
Cortisol ± 15 nmol/L < 100 nmol/L ± 15% > 100 nmol/L
Creatinine ± 8 umol/L < 100 umol/L ± 8% > 100 umol/L
Ferritin ± 4.0 ug/L < 27.0 ug/L ± 15% > 27 ug/L
Fructosamine ± 15 umol/L < 250 umol/L ± 6% > 250 umol/L
Glucose ± 0.4 mmol/L < 5.0 mmol/L ± 8% > 5.0 mmol/L
GGT ± 5 U/L < 40 U/L ± 12% > 40 U/L
hCG-quantitative ± 1.0 IU/L < 10 IU/L ± 10% > 10 IU/L
hCG-qualtitative (pre 2012) neg < 20 IU/L pos > 30 IU/L
HDL cholesterol ± 0.1 mmol/L < 0.8 mmol/L ± 12% > 0.8 mmol/L
Iron ± 3 umol/L < 25 umol/L ± 12% > 25 umolL
Lactate Dehydrogenase ± 20 U/L < 250 U/L ± 8% > 250 U/L
Lactate ± 0.5 mmol/L < 4 mmol/L ± 12% > 4 mmol/L
Lipase ± 12 U/L < 60 U/L ± 20% > 60 U/L
Lithium ± 0.2 mmol/L < 2.0 mmol/L ± 10% > 2.0 mmol/L
Magnesium
± 0.1 mmol/L < 1.25 mmol/L
± 8% > 1.25 mmol/L
Osmolality ± 8 mmol/kg < 266 mmol/kg ± 3% > 266 mmol/kg
Phosphate ± 0.06 mmol/L < 0.75 mmol/L ± 8% > 0.75 mmol/L
Potassium ± 0.2 mmol/L < 4.0 mmol/L ± 5% > 4.0 mmol/L
Protein (Total) ± 3.0 g/L < 60 g/L ± 5% > 60 g/L
Sodium ± 3 mmol/L < 150 mmol/L ± 2% > 150 mmol/L
Total T3 ± 0.2 nmol/L < 1.3 nmol/L ± 15% > 1.3` nmol/L
Free T3 ± 0.7 pmol/L < 3.5 pmol/L ± 20% > 3.5 pmol/L
Free T4 ± 1.5 pmol/L < 12 pmol/L ± 12% > 12 pmol/L
Total T4 ± 12 nmol/L < 120 nmol/L ± 10% > 120` nmol/L
TSH ± 0.1 mU/L < 0.5 mU/L ± 20% > 0.5 mU/L
Thyroxine ± 12 nmol/L < 120 nmol/L ± 10% > 120 nmol/L
TIBC ± 4 umol/L < 50 umol/L ± 8% > 50 umol/L
Transferrin ± 0.2 g/L < 2.5 g/L ± 8% > 2.5 g/L
Triglyceride ± 0.2 mmol/L < 1.6 mmol/L ± 12% > 1.6 mmol/L
Troponin I ± 0.002 ug/L < 0.10 ug/L ± 20% > 0.10 ug/L
Troponin T ± 0.01 ug/L < 0.05 ug/L ± 20% > 0.05 ug/L
Urate ± 0.03 mmol/L < 0.38 mmol/L ± 8% > 0.38 mmol/L
Urea ± 0.5 mmol/L < 4.0 mmol/L ± 12% > 4.0 mmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

General Therapeutic Drugs - updated 2012
Test or Analyte Lower Goal Upper Goal
Carbamazepine ± 0.5 mg/L < 4.7 mg/L ± 10% > 4.7 mg/L
Digoxin ± 0.2 ug/L < 1.6 ug/L ± 10% > 1.6 ug/L
Gentamicin ± 0.2 mg/L < 2.0 mg/L ± 10% > 2.0 mg/L
Paracetamol ± 3.0 mg/L < 30 mg/L ± 10% > 30 mg/L
Phenobarbitone ± 0.7 mg/L < 7.0 mg/L ± 10% > 7.0 mg/L
Phenytoin ± 8.0 mg/L < 7.6 mg/L ± 10% > 7.6 mg/L
Quinidine (pre 2012) ± 2.0 umol/L < 20.0 umol/L ± 10% > 20.0 umol/L
Salicylate ± 14.0 mg/L < 143.0 mg/L ± 10% > 143 mg/L
Theophylline ± 0.5 mg/L < 5.4 mg/L ± 10% > 5.4 mg/L
Valproate ± 4 mg/L < 36 mg/L ± 10% > 36 mg/L
Vancomycin ± 0.2 mg/L < 20.3 mg/L ± 10% > 20.3 mg/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Glycohaemoglobin - updated 2013
Test or Analyte Lower Goal Upper Goal
HbA1c (NGSP) (%) ± 0.4 < 6.7% ± 6% > 6.7%
HbA1c (IFCC) (mmol/mol) ± 4 mmol/mol < 45 mmol/mol ± 8% > 45 mmol/mol

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

IGF-1 / C Peptide
Test or Analyte Lower Goal Upper Goal
IGF-1 ± 3 nmol/L < 25.0 nmol/L ± 12% > 25.0 nmol/L
C Peptide ± 0.15 nmol/L < 1.25 nmol/L ± 12% > 1.25 nmol/L
IGFBP-3 ± 15.0 nmol/L < 100 nmol/L ± 15% > 100 nmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Near Patient Testing - updated 2012
Test or Analyte Lower Goal Upper Goal
Albumin ± 2.0 g/L < 33.0 g/L ± 6% > 33.0 g/L
Alkaline Phosphatase ± 15 U/L < 125 U/L ± 12% > 125 U/L
ALT ± 5 U/L < 40 U/L ± 12% > 40 U/L
Amylase ± 10 U/L < 100 U/L ± 10% > 100 U/L
AST ± 5 U/L < 40 U/L ± 12% > 40 U/L
Bicarbonate ± 2.0 mmol/L < 20.0 mmol/L ± 10% > 20.0 mmol/L
Bilirubin Conjugated ± 3.0 umol/L < 15.0 umol/L ± 20% > 15.0 umol/L
Bilirubin Total ± 3.0 umol/L < 25.0 umol/L ± 12% > 25.0 umol/L
Calcium ± 0.1 mmol/L < 2.5 mmol/L ± 4% > 2.5 mmol/L
Chloride ± 3 mmol/L < 100 mmol/L ± 3% > 100 mmol/L
Cholesterol ± 0.30 mmol/L < 5.0 mmol/L ± 6% > 5.0 mmol/L
Creatinine Kinase ± 15 U/L < 125 U/L ± 12% > 125 U/L
Creatinine ± 8.0 umol/L < 100 umol/L ± 8% > 100 umol/L
Glucose ± 0.4 mmol/L < 5.0 mmol/L ± 8% > 5.0 mmol/L
Creatinine ± 0.01 mmol/L < 0.1 mmol/L ± 10% > 5.0 mmol/L
GGT ± 5 U/L < 40 U/L ± 12% > 40 U/L
hCG-quantitative (pre 2012) ± 3 IU/L < 20 IU/L ± 15% > 20 IU/L
hCG-qualitative (pre 2012) neg < 20 IU/L pos > 30 IU/L
HDL Cholesterol ± 0.1 mmol/L < 0.8 mmol/L ± 12% > 0.8 mmol/L
Lactate Dehydrogenase ± 20 U/L < 250 U/L ± 8% > 250 U/L
Magnesium ± 0.1 mmol/L < 1.25 mmol/L ± 8% > 1.25 mmol/L
Phosphate ± 0.06 mmol/L < 0.75 mmol/L ± 8% > 0.75 mmol/L
Potassium ± 0.2 mmol/L < 4.0 mmol/L ± 5% > 4.0 mmol/L
Protein, Total ± 3 g/L < 60 g/L ± 5% > 60 g/L
Sodium ± 3 mmol/L < 150 mmol/L ± 2% > 150 mmol/L
Triglyceride ± 0.2 mmol/L < 1.6 mmol/L ± 12% > 1.6 mmol/L
Urate ± 0.03 mmol/L < 0.38 mmol/L ± 8% > 0.38 mmol/L
Urea ± 0.5 mmol/L < 4.0 mmol/L ± 12% > 4.0 mmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Neonatal Bilirubin - updated 2012
Test or Analyte Lower Goal Upper Goal
Total Bilirubin ± 8 umol/L < 80 umol/L ± 10% > 80 umol/L
Conjugated Bilirubin ± 3 umol/L < 15 umol/L ± 20% > 15 umol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Porphyrins
Test or Analyte Lower Goal Upper Goal
Urine ALA ± 8 umol/L < 32 umol/L ± 25% > 32 umol/L
Urine Porphobilinogen
(Quantitative)
± 5 umol/L < 20 umol/L ± 25% > 20 umol/L
Urine Porphobilinogen
Qualitative (pre 2012)
neg < 10 umol/L + 10-50 umol/L
++ 51-100 umol/L
+++ > 100 umol/L
Urine Total Porphyrin ± 60 nmol/L < 300 nmol/L ± 20% > 300 nmol/L
Faecal Percent Dry Weight ± 2 < 20% ± 10% > 20%
Faecal Total Porphyrin ± 40 umol/kg < 200 umol/kg ± 20% > 200 umol/kg
Plasma Total Porphyrin ± 4 nmol/L < 20 nmol/L ± 20% > 20 nmol/L
RBC Total Porphyrin ± 0.4 umol/L rc < 2.0 umol/L rc ± 20% > 2.0 umol/L rc

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Special lipids - updated 2012
Test or Analyte Lower Goal Upper Goal
Cholesterol ± 0.30 mmol/L < 5.0 mmol/L ± 6% > 5.0 mmol/L
HDL Cholesterol ± 0.10 mmol/L < 0.8 mmol/L ± 12% > 0.8 mmol/L
LDL Cholesterol ± 0.20 mmol/L < 2.0 mmol/L ± 10% > 2.0 mmol/L
Triglyceride ± 0.20 mmol/L < 1.6 mmol/L ± 12% > 1.6 mmol/L
Apolipoprotein A1 ± 0.2 g/L < 2.0 g/L ± 10% > 2.0 g/L
Apolipoprotein B ± 0.2 g/L < 2.0 g/L ± 10% > 2.0 g/L
Lipoprotein(a) ± 0.2 g/L < 2.0 g/L ± 10% > 2.0 g/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Special Therapeutic Drugs - updated 2012
Test or Analyte Lower Goal Upper Goal
Amiodarone ± 0.1 mg/L < 1.3 mg/L ± 10% > 1.3 mg/L
Amitriptyline ± 3.0 ug/L < 28.0 ug/L ± 10% > 28.0 ug/L
Clozapine ± 10 ug/L < 100 ug/L ± 10% >100 ug/L
Disopyramide (pre 2012) ± 2.0 umol/L < 20 umol/L ± 10% > 20 umol/L
Ethosuximide (pre 2012) ± 3 umol/L < 30 umol/L ± 10% > 30 umol/L
Lignocaine (pre 2012) ± 2.0 umol/L < 20 umol/L ± 10% > 20 umol/L
Methotrexate ± 0.1 umol/L < 1.0 umol/L ± 10% > 1.0 umol/L
Mycophenylate ± 0.1 mg/L < 1.0 mg/L ± 10% > 1.0 mg/L
Nortriptyline ± 3.0 ug/L < 26.0 ug/L ± 10% > 26.0 ug/L
Quinidine (pre 2012) ± 2.0 umol/L < 20 umol/L ± 10% > 20 umol/L
Tricyclic Antidepressants (pre 2012) ± 10 nmol/L < 100 nmol/L ± 10% > 100 nmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Sweat Electrolytes - updated 2013
Test or Analyte Lower Goal Upper Goal
Sodium ± 2 mmol/L < 40 mmol/L ± 5% > 40 mmol/L
Chloride ± 2 mmol/L < 40 mmol/L ± 5% > 40 mmol/L
Sweat Conductivity ± 2 mmol/L < 50 mmol/L ± 5% > 50 mmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Tumour Markers - updated 2012
Test or Analyte Lower Goal Upper Goal
ACTH ± 2.0 pmol/L; < 20 pmol/L ± 10%; > 20 pmol/L
Alpha-fetoprotein ± 2 kIU/L; < 17 kIU/L ± 12% > 17 kIU/L
Calcitonin ± 2 ng/L < 20 ng/L ± 10% > 20 ng/L
CEA ± 0.6 ug/L < 5.0 ug/L ± 12% > 5.0 ug/L
CA125 ± 6 kU/L < 50 kU/L ± 12% > 50 kU/L
CA15-3 ± 3 kU/L < 30 kU/L ± 10% > 30 kU/L
CA19-9 ± 6 kU/L < 40 kU/L ± 15% > 40 kU/L
hCG ± 1 IU/L < 10 IU/L ± 10% > 10 IU/L
Beta-2 Microglobulin ± 0.2 mg/L < 2.0 mg/L ± 10% > 2.0 mg/L
NSE ± 2.0 ug/L < 20 ug/L ± 10% > 20 ug/L
Gastrin ± 10 pmol/L < 100 pmol/L ± 10% > 100 pmol/L
Prolactin ± 40 mIU/L < 400 mIU/L ± 10% > 400 mIU/L
Total PSA ± 0.4 ug/L < 5.0 ug/L ± 8% > 5.0 ug/L
Thyroglobulin ± 0.2 ug/L < 1.7 ug/L ± 12% > 1.7 ug/L
P1NP ± 5 ug/L < 50 ug/L ± 10% > 50 ug/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Urine Chemistry
Test or Analyte Lower Goal Upper Goal
Albumin ± 0.1 g/L < 1.0 g/L ± 10% > 1.0 g/L
Calcium
± 0.20 mmol/L
Chloride ± 2.0 mmol/L < 20.0 mmol/L ± 10% > 20.0 mmol/L
Creatinine ± 0.5 mmol/L < 5.0 mmol/L ± 10% > 5.0 mmol/L
DPD (Total) ± 15 nmol/L < 80 nmol/L ± 15% > 80 nmol/L
DPD (Free) ± 15 nmol/L < 80 nmol/L ± 15% > 80 nmol/L
Glucose ± 1.0 mmol/L < 10.0 mmol/L ± 10% > 10.0 mmol/L
hCG-qualitative neg < 20 IU/L pos > 30 IU/L
hCG-quantitative ± 5 IU/L < 50 IU/L ± 10%; > 50 IU/L
Hydroxyproline ± 10 umol/L < 100 umol/L ± 10%; > 100 umol/L
Magnesium (serum)
± 0.12 mmol/L
N-Telopeptides ± 80 nmolBCE/L < 700 nmolBCE/L ± 10% > 700 nmolBCE/L
Osmolality ± 6 mmol/kg < 300 mmol/kg ± 2% > 300 mmol/kg
Potassium ± 2.0 mmol/L < 20.0 mmol/L ± 10% > 20.0 mmol/L
Phosphate
± 2.5 mmol/L
Sodium ± 2.0 mmol/L < 20.0 mmol/L ± 10% > 20.0 mmol/L
Total Protein ± 0.1 g/L < 1.0 g/L ± 10% > 1.0 g/L
Urate
± 0.3 mmol/L
Urea ± 20 mmol/L < 200 mmol/L ± 10% > 200 mmol/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Urine Pregnancy Testing
Test or Analyte Lower Goal Upper Goal
hCG neg < 20 IU/L pos > 30 IU/L

Back to top


Note: Analytical Quality Requirements are a plus/minus percentage. Some browsers may not display the plus/minus character (it should appear right here, ±. If it does not, your browser does not support this character).

Vitamins
Test or Analyte Lower Goal Upper Goal
Vitamin A ± 0.3 umol/L < 3.0 umol/L ± 10% > 3.0 umol/L
Vitamin E ± 5.0 umol/L < 50.0 umol/L ± 10% > 50.0 umol/L
Beta Carotene ± 0.2 umol/L < 20.0 umol/L ± 10% > 20.0 umol/L
Total Carotenoids ± 0.4 umol/L < 4.0 umol/L ± 10% > 4.0 umol/L