Tools, Technologies and Training for Healthcare Laboratories

QC Selection Grid, Square 4

MULTIRULE CHOICE:

Mqcsg4

This is not a good situation because you have a moderate number of problems and medically important errors are small and difficult to detect. You should use a five rule procedure with N=4. The 8x rule should be used across runs and interpreted as a rejection rule. The power curves below show the rejection characteristics within a run (R=1) and across two runs (R=2):

Mpf2

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curves. Where the curves and critical-error meet indicates how much error detection each control rule will provide.

Let us know what you're interested in!

Please use this form to request more information about.

Westgard Products and Services.

Invalid Input
Invalid Input
Invalid Input
Invalid Input