Posted by Sten Westgard, MS

In our first webinar, we had a lot of questions come up - thanks to all the participants who submitted them.

I'm going to address them in the order in which they were submitted, so I apologize for the random order.

From Jakarta: "How do we draw OPSpecs charts? By Excel?"

OPSpecs charts can be downloaded directly from the Westgard Website. It requires fairly sophisticated knowledge to build them, so we usually don't see them done by Excel. There are a few programs that build dynamic OPSpecs charts, including the Bio-Rad Westgard Advisor, and our own EZ Rules 3.

From Hyderabad: "What is the exact IQC run frequency in major hospital laboratories?"

This is an interesting question. We can answer it in two ways. There is the QC frequency that current laboratories are performing, and that's usually once a day. This frequency is based more on compliance and regulatory minimums, and has little relationship to the needs of the patient. But the more interesting question is what QC frequency should major laboratories be performing? That answer is based on volume and performance of the assays. It could be that an assay with high quality can still run only once a day even in high volume labortories, but the forthcoming QC frequency tools will soon reveal that a low quality assay may need to run controls every 100 patients or even every 10 patients!

From Tamluk: "Please comment on the 10x rule"

This is a very common question - often because laboratories are seeing a 10:x violation and wonder if they truly need to reject their run(s). The short answer is, if the Sigma-metric is 4 or higher, the 10:x rule is not a necessary rejection rule. Only when performance falls below 4 Sigma do the 10:x (or 6:x. 8:x, 9:x, 10:x, or 12:x rules) become required for proper error detection.
Another tip on using the 10:x rule - make sure there aren't some rounding issues causing the values to fall above or below the mean. If you have a "mean" of 10.5 on a test where your system does not report any decimal places, to give a simplistic example, then any value 10 or 11 is going to be above or below the mean. If you suspect the rounding of your system is causing too many values to fall above or below the mean, try to increase the number of decimal places reported in your system. Internally, you can always use an extra decimal place, although you won't report that extra decimal place in your results.

From Kolkata: "Are "Westgard Rules" only applicable for the diagnostics field?"

We rarely see "Westgard Rules" used outside the diagnostic laboratory medicine field, partly because in other fields, it's pretty cheap to run additional controls. One of the primary reasons we use multirule QC procedures is the expense of controls - we can't afford to run many of them, so we need to get as many rule interpretations out of each measurement as possible. But we have anecdotally heard that rocket engine testing and some anthropological testing (both cases where the number of measurements is small), have implemented "Westgard Rules" in their QC practices.

From Kolkata: "What is the frequency for running 2 levels of control if a lab is processing more than 400 samples a day?"

Like the question above, this has two answers. For compliance purposes, most countries will be satisfied with QC run once a day (of course for hematology and coag and some other fields, a higher frequency, perhaps 3 times per day, is required). But the scientific answer to this question will require more information about the quality required by the test (TEa) and the performance observed (CV, bias) of the method. If we know this is a Six Sigma test, for example, we only need to run QC once a day and we could use just about any QC rule. But a 3 Sigma method will require more frequent QC.

From Indonesia: "What is the best way to troubleshoot when we are out of control?"

There are a lot of things you should NOT do when you're out-of-control, such as repeating controls without any investigation, or running new controls without any trouble-shooting. There are times when you do want to repeat the control (to confirm you are back in-control, for instance), but you should be trouble-shooting and seeking the root cause of the type of error (random or systematic). There are a lot of different trouble-shooting flowcharts you can consult, including some from our Basic QC Practices manual.

From India: "What is your suggestion when there is a narrow or very narrow SD which leads to false rejection?"

This is a situation that occurs all too frequently. We call it "anorexic QC", where the continuous use of tight 2 SD limits, combined with continuous exclusion of 2 SD outliers, can progressively narrow the limits.  The answer is to move away from the 2 SD limits toward a wider QC design (assuming the Sigma-metric justifies it). It could be that over a longer period of time, the use of wider limits will result in a healthier SD (it will lose its anorexia) and then the QC design will be adjusted again to something less wide.  Every few months you should look at performance, and it if has significantly changed (for example, shifting a Sigma-metric from 4 to 5, etc.), you need to re-design your QC procedures.
I usually see laboratories in anorexic QC situations, but rarely do I see them "fatten" up. It would be interesting if a laboratory wanted to share this journey with us - I think it would actually be something that could be published.

From Kolkata: "Can replicate testing in a laboratory check the IQC of the test results?"

I'm having a little bit of a problem figuring out what this question is really trying to ask. Replicate testing of the patient samples is a last resort method of improving performance. If you have a method with a Sigma-metric below 3, it's something you may need to consider. By testing twice, and using that to generate just one results, you reduce the variation in the answer. But it's the equivalent of doubling your test volume, a very expensive step. Now, if the laboratory is actually suggesting making replicate measurements of the control, that is the equivalent of doubling your control measurements. So if you run two controls, but you measure them twice, that gives you effectively 4 control measurements to use in your QC charts. So this is a way to increase your QC without spending more on controls.

From Tamluk: "Is it mandatory to run every level of controls of each test? Like there are up to 4 levels of control for HIV? Is there a requirement to run every level every day?"

The answer to this will depend on the regulations that apply in your country, as well as the manufacturer recommendations. If the regulations only require QC once a day but only a positive and negative, you may be free to run fewer controls than are available. But if the manufacturer instructions recommend that you run 4 levels of control, you really should run those 4 levels. In the US, if you violate the manufacturer recommendations, you have effectively modified the test into a high complexity method and there are a huge number of new method validation studies you must perform. So the short answer is to do what the manufacturer and regulator recommend. Then and only then, can you take into account the actual performance of the method. If your regulations and manufacturer don't specify the number of controls to run, and you have a Six Sigma method, you may consider going to just two controls per run (one low positive and one negative).

From Hyderabad: "Is 1:2s acceptable in hematology?"

Here I must use the dreaded response, "It depends."  If you are performing your QC manually, using your own brain power, and charts drawn on paper, then a 1:2s warning rule may be useful. If you have any kind of software performing your QC, we recommend using only rejection rules and skipping the 1:2s rule as a warning rule. Note this is true for any type of testing, not just hematology or chemistry. The 1:2s rule generates a lot of false rejections, and you don't want to spend your precious time chasing those ghosts.

There were more than 50 questions submitted during the first webinar. We will get to more questions in another post.

-----