MULTIRULE CHOICE:

Mqcsg5

You have a moderate number of problems and medically important errors are moderate in size (errors neither hard nor easy to detect). It is recommended that you increase the number of rules to four and apply the 41s rule across two runs (R=2) as a rejection rule. The power curves below show the rejection characteristics within one run (R=1) and across two runs (R=2):


Mpf3

Plot your critical-error on the x-axis of this power function graph. Draw a vertical line from that point to intersect the power curves. Where the curves and critical-error meet indicates how much error detection each control rule will provide.